DosimetryDoseCalcOpt
DoseCalcOpt

The Dose Calc. Options menu controls which factors are applied to the dose calculation. YOU SHOULD VERIFY THE SETTINGS IN THIS MENU PRIOR TO PERFORMING A DOSE CALCULATION !!! The affected factors are anisotropy, linear vs point source, a correction for plaques with silicone carriers, a scatter correction, lip collimation, slot collimation, and dose constants.

  • COMS calculation - forces dose calculation to use the COMS formalism. Disabled by default.
  • Anisotropy F(φ) - enables source anisotropy. Enabled by default for plaques with source slots.
  • Linear sources - Enables linear sources. Enabled by default for plaques with source slots.
  • Carrier Attenuation - enable correction for COMS plaque carriers. Enabled by default for COMS plaques.
  • Scatter - this hierarchical menu accesses dosimetric corrections which attempt to compensate for the lack of a homogeneous full scatter geometry.
  • Near Field - .
  • Shell collimation - enables collimation of primary radiation which passes through the plaque shell. Enabled by default.
  • Slot collimation - collimates primary radiation which passes outside of slots in the plaque surface. Disabled by default.
  • Partial exposure - Calculates fraction of primary radiation from a linear source which does not pass through the shell. Enabled by default.
  • Penumbra - Enables estimation of penumbral edges. Disabled by default.

Dosimetry menu | Rx Window | Guide Contents